ΑΓΩΓΙΜΟΜΕΤΡΟ ΕΡΓΑΣΤΗΡΙΑΚΟ, ΜΕ ΚΑΤΑΓΡΑΦΗ (EC, TDS, NaCl and Temperature)

624,94  με ΦΠΑ

MW170 4-in-1 EC/TDS/NaCl/Temperature Logging bench meter.

Range: 0.00 up to 500.0 mS/cm, up to 250.0 g/L absolute TDS, 0.0 to 400.0 % NaCl-20.0 to 120.0 ºC.

Data logging: 1000 logs can be stored in the built-in memory including readings.

It is accurate to ± 1% of reading for EC/TDS & Salinity with automatic, manual or no temperature compensation and a built in dedicated GLP key.

Ideal for use by education, horticulture, labs, food processing, water treatment and many other applications with challenging EC/TDS/NaCL/Temp. measurement needs.

Supplied with MA814DB/1 EC/TDS/NaCl/Temperature probe, electrode holder and adapter.

Σε απόθεμα

Κωδικός προϊόντος: 64212 Κατηγορίες: ,

Features

  • Fast easy to read results with large duel level digital LCD readout.
  • Extreme accuracy to 1% of reading for EC/TDS/NaCl |+/- 0.5°C (+/-0.9°F). 
  • Extended auto-ranging feature for EC/TDS/NaCl.
  • Lightweight, sleek design featuring a small benchtop footprint.
  • Simple easy-to-use logging; manual log-on-demand or manual log-on-stability up to 1,000 records.  
  • Micro USB Port download functions. 
  • GLP data, date and time.
  • Easy measurement: just place the probe in the sample, gently stir and wait for the reading to stabilize.
  • Single cell factor calibration for EC/TDS/NaCl, automatic simple one point calibration and two point for temperature allowing increased accuracy.
  • Rugged low maintenance, high performance, easy-to-clean EC/TDS/NaCL/Temp probe.
  • Built in temperature sensor. 
  • Internal clock date to keep track of different time-dependent functions (calibration, timestamp,calibration time out).
  • 3 feet probe cables for extra flexibility when testing.
  • Includes MA9315 electrode holder, cables, adapter and probe.

Care, Use and Cleaning

EC meters usually start performing poorly because of problems with the sediment on the conductivity probe’s rings. The dirty rings on the probe can change impedance. Cleaning the rings is simple.  Slide the plastic shield off and clean with DI water and soft cloth to keep your readings accurate.

Storage

When not in use, store the meter and probe in a safe area to avoid damage. The EC probe is stored dry. 

Calibration

The MW170 MAX is easy 1 point EC/TDS/NaCl calibration for ease-of-use and two point temperature calibration for best accuracy in (ATC).  Calibrating regularly is best to keep your PRO MW170 4-in-1 MAX Bench meter accurate for all your readings.